Abstract
Today, visual classification of the degree of lipemia in blood samples is frequently performed in clinical laboratories. However, achieving standardization of this classification at low cost and with fewer resources is an objective that is still under development. In this work, a comparison is made between the visual inspection and optical measurements of blood plasma for quantifying lipemia. The plasma refractive index’s real part was measured using an Abbe refractometer and transmittance measurements were made at a 589 nm wavelength and wavelengths ranging from 320 to 1100 nm in the spectral region, respectively. Taking the slope of the transmittance spectrum at two specific wavelengths, it is possible to establish a more standardized selection criterion and implement it quickly using low-cost optical devices. Furthermore, using the proposed transmittance-spectrum-slope method, statistically significant differences (p < 0.05) were found between healthy blood samples and lipemia 1, 2, 3, and 4. However, there were no statistical differences between lipemia 1 and 2.