Evaluation of the Abbott Alinity i Thyroid-Stimulating Hormone Receptor Antibody (TRAb) Chemiluminescent Microparticle Immunoassay (CMIA)

Author:

Lee Deborah J. W.1,Phua Soon Kieng1,Liang Yali1,Chen Claire2,Aw Tar-Choon134ORCID

Affiliation:

1. Department of Laboratory Medicine, Changi General Hospital, Singapore 529889, Singapore

2. Abbott Laboratories, Singapore 189352, Singapore

3. Yong Loo Lin School of Medicine, National University of Singapore (NUS), Singapore 119228, Singapore

4. Duke-NUS Graduate School of Medicine, Singapore 169857, Singapore

Abstract

Background: We evaluated the performance of the Abbott thyroid-stimulating hormone receptor antibody chemiluminescent microparticle immunoassay (CMIA) on the Alinity i. Methods: Verification studies for precision, linearity, analytical measuring range, diagnostic cut offs for Graves’ disease were performed. We compared the Abbott CMIA to an established TRAb assay (Roche electrochemiluminescence immunoassay). Method comparison analysis was performed between serum and plasma samples on the Abbott CMIA. Results: Repeatability (CV%) for TRAb were 4.07, 1.56, 0.71 and within-laboratory imprecision (CV%) were 4.07, 1.90, 0.71 at 3.0, 10.0, 30.0 IU/L of TRAb, respectively. Linearity and analytical measuring range were verified from 1.07–47.9 IU/L. The limit of the blank was 0 IU/L, limit of detection was 0.15 IU/L, and limit of quantification was 0.5 IU/L. Passing-Bablok analysis showed agreement between the two assays; Y-intercept = 0.787, slope = 1.04. Passing-Bablok analysis also showed agreement between the plasma and serum samples run on the Abbott CMIA; Y-intercept −0.17, slope = 0.97. Conclusions: The Abbott TRAb CMIA on the Alinity i performs within the manufacturer claims for assay precision, linearity, analytical measuring range, limit of blank, limit of detection, limit of quantitation and diagnostic cut offs for Graves’ disease. Thus, the Abbott TRAb CMIA on the Alinity i is fit for clinical use.

Funder

Abbott Diagnostics

Publisher

MDPI AG

Subject

Clinical Biochemistry

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3