Abstract
The instability of human serum albumin (HSA) in urine samples makes fresh urine a requirement for microalbumin analyses using immunoturbidimetry. Here, we determined the ability of an aptasensor-based fluorescent platform to detect microalbumin in old, boric acid-preserved urine samples. Our results show that the cleavage site of protease enzymes on urine albumin protein differed from the binding position of the aptamer on HSA protein, suggesting the aptasensor may be effective for albumin detection in non-fresh urine. Furthermore, the addition of boric acid in urine samples over a short term (at ambient temperature (Ta) and 4 °C), long term (−20 and −80 °C), and following freeze–thawing (1–3 cycles) did not significantly affect albumin stability, as analyzed using the aptasensor. Therefore, boric acid stabilized has in urine stored over a short- and long-term. Thus, the aptasensor developed by us is applicable for HSA detection in boric acid-preserved urine that has been stored for 7-d at Ta and 4 °C, and in the long-term at −80 °C.
Funder
National Nanotechnology Center (NANOTEC), National Science and Technology Development Agency (NSTDA), Thailand
Program Management Unit B (PMUB), Thailand Science Research and Innovation
Subject
Clinical Biochemistry,General Medicine,Analytical Chemistry,Biotechnology,Instrumentation,Biomedical Engineering,Engineering (miscellaneous)
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献