Abstract
An innovative electrochemical biosensor based on graphene quantum dots (GQDs) is developed for a simple, rapid, and highly sensitive primary diagnosis of the breast cancer biomarker cluster of differentiation-44 (CD44) antigen. Herein, electrochemical exfoliation of waste dry batteries provides facile, eco-friendly, and cost-effective synthesis of GQDs. Transmission electron microscopy (TEM) analysis reveals that GQDs exhibit spherical shapes with an average diameter of 4.75 nm. Further, electrochemical analysis through cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) reveals that the electrochemical properties of GQDs are suitable for biosensing applications. Subsequently, GQDs have a large electroactive surface area that has been utilized for the immobilization of CD44 antibodies to fabricate the electrochemical biosensor. The electroanalytical performance of GQDs for CD44 biosensing capabilities is studied by differential pulse voltammetry (DPV). The developed electrochemical biosensor has high sensitivity with the lowest detection limit (LOD) of 2.11 fg/mL in the linear range of 0.1 pg/mL to 100.0 ng/mL in phosphate buffer saline (PBS). Further, the linear response of the electrochemical biosensor for CD44 antigen concentration is in the range of 1.0 pg/mL to 100.0 ng/mL with a LOD of 2.71 fg/mL in spiked serum samples. The outcomes suggest that the synthesized GQDs demonstrate promising attributes to be utilized as a viable nanomaterial in biosensing applications.
Subject
Clinical Biochemistry,General Medicine,Analytical Chemistry,Biotechnology,Instrumentation,Biomedical Engineering,Engineering (miscellaneous)
Cited by
18 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献