Photoneutrons and Gamma Capture Dose Rates at the Maze Entrance of Varian TrueBeam and Elekta Versa HD Medical Linear Accelerators

Author:

Suliman Ibrahim I.ORCID,Khouqeer Ghada A.ORCID,Mayhoub Fareed H.

Abstract

Herein, we evaluated the neutron and gamma capture dose equivalent rates at the maze entrance of Varian TrueBeam and Elekta Versa HD™ medical linear accelerators (linacs) using experimental measurements as well as empirical calculations. Dose rates were measured using calibrated neutron and gamma area survey meters placed side-by-side at the measurement point of interest. Measurements were performed at a source-to-detector distance of 100 cm, with a 10 × 10 cm2 field size therapeutic X-ray beam, and a 30 × 30 × 15 cm3 solid water patient equivalent phantom, with a linac operating at 15, 10 MV, and 10 MV flattened filter-free (FFF). Dose rates were also measured at different points at the centerline along the maze towards the maze entrance. The measured dose equivalent rates at the maze entrance were comparable to those reported in the literature. The dose rates along the maze decreased exponentially towards the maze entrance and were significant for short maze lengths. The evaluated empirical methods for estimating neutron dose rates at the maze entrance of a linac proposed by Kersey, the modified Kersey method and Falcão method, agree by a factor of two from the experimental measurements. The results revealed vital radiation protection considerations owing to neutron contamination in external beam therapy.

Funder

Deanship of Scientific Research, Imam Mohammad Ibn Saud Islamic University

Publisher

MDPI AG

Subject

Chemical Health and Safety,Health, Toxicology and Mutagenesis,Toxicology

Reference35 articles.

1. Global cancer incidence and mortality rates and trends—An update;Torre;Cancer Epidemiol. Biomark. Prev.,2016

2. Neutron contamination in radiotherapy processes: A review study;Banaee;J. Radiat. Res.,2021

3. Neutron spectra and dosimetric features around an 18 MV linac accelerator;Barquero;Health Phys.,2005

4. Measurement of photo-neutron dose from an 18-MV medical linac using a foil activation method in view of radiation protection of patients;Kaya;Nucl. Eng. Technol.,2016

5. Measurements of secondary neutron dose from 15 MV and 18 MV IMRT;Howell;Radiat. Prot. Dosim.,2005

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献