Abstract
The Oticon Medical Neuro cochlear implant system includes the modes Opti Omni and Speech Omni, the latter providing beamforming (i.e., directional selectivity) in the high frequencies. Two studies compared sentence identification scores of adult cochlear implant users with Opti Omni and Speech Omni. In Study 1, a double-blind longitudinal crossover study, 12 new users trialed Opti Omni or Speech Omni (random allocation) for three months, and their sentence identification in quiet and noise (+10 dB signal-to-noise ratio) with the trialed mode were measured. The same procedure was repeated for the second mode. In Study 2, a single-blind study, 11 experienced users performed a speech identification task in quiet and at relative signal-to-noise ratios ranging from −3 to +18 dB with Opti Omni and Speech Omni. The Study 1 scores in quiet and in noise were significantly better with Speech Omni than with Opti Omni. Study 2 scores were significantly better with Speech Omni than with Opti Omni at +6 and +9 dB signal-to-noise ratios. Beamforming in the high frequencies, as implemented in Speech Omni, leads to improved speech identification in medium levels of background noise, where cochlear implant users spend most of their day.
Funder
Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献