Dual and Multi-Target Cone-Beam X-ray Luminescence Computed Tomography Based on the DeepCB-XLCT Network
-
Published:2024-08-28
Issue:9
Volume:11
Page:874
-
ISSN:2306-5354
-
Container-title:Bioengineering
-
language:en
-
Short-container-title:Bioengineering
Author:
Liu Tianshuai12, Huang Shien13, Li Ruijing12, Gao Peng12, Li Wangyang12, Lu Hongbing12ORCID, Song Yonghong3ORCID, Rong Junyan12ORCID
Affiliation:
1. Biomedical Engineering Department, Fourth Military Medical University, Xi’an 710032, China 2. Shaanxi Provincial Key Laboratory of Bioelectromagnetic Detection and Intelligent Perception, Xi’an 710032, China 3. School of Software Engineering, Xi’an Jiaotong University, Xi’an 710049, China
Abstract
Background and Objective: Emerging as a hybrid imaging modality, cone-beam X-ray luminescence computed tomography (CB-XLCT) has been developed using X-ray-excitable nanoparticles. In contrast to conventional bio-optical imaging techniques like bioluminescence tomography (BLT) and fluorescence molecular tomography (FMT), CB-XLCT offers the advantage of greater imaging depth while significantly reducing interference from autofluorescence and background fluorescence, owing to its utilization of X-ray-excited nanoparticles. However, due to the intricate excitation process and extensive light scattering within biological tissues, the inverse problem of CB-XLCT is fundamentally ill-conditioned. Methods: An end-to-end three-dimensional deep encoder-decoder network, termed DeepCB-XLCT, is introduced to improve the quality of CB-XLCT reconstructions. This network directly establishes a nonlinear mapping between the distribution of internal X-ray-excitable nanoparticles and the corresponding boundary fluorescent signals. To improve the fidelity of target shape restoration, the structural similarity loss (SSIM) was incorporated into the objective function of the DeepCB-XLCT network. Additionally, a loss term specifically for target regions was introduced to improve the network’s emphasis on the areas of interest. As a result, the inaccuracies in reconstruction caused by the simplified linear model used in conventional methods can be effectively minimized by the proposed DeepCB-XLCT method. Results and Conclusions: Numerical simulations, phantom experiments, and in vivo experiments with two targets were performed, revealing that the DeepCB-XLCT network enhances reconstruction accuracy regarding contrast-to-noise ratio and shape similarity when compared to traditional methods. In addition, the findings from the XLCT tomographic images involving three targets demonstrate its potential for multi-target CB-XLCT imaging.
Funder
Natural National Science Foundation of China National Key Research and Development Program Key Research and Development Program of Shaanxi Province
Reference35 articles.
1. Hybrid x-ray/optical luminescence imaging: Characterization of experimental conditions;Carpenter;Med. Phys.,2010 2. Liu, T., Ruan, J., Rong, J., Hao, W., Li, W., Li, R., Zhan, Y., and Lu, H. (2022). Cone-beam X-ray luminescence computed tomography based on MLEM with adaptive FISTA initial image. Comput. Methods Programs Biomed., 229. 3. Zhang, H., Hai, L., Kou, J., Hou, Y., He, X., Zhou, M., and Geng, G. (2022). OPK_SNCA: Optimized prior knowledge via sparse non-convex approach for cone-beam X-ray luminescence computed tomography imaging. Comput. Methods Programs Biomed., 215. 4. Lun, M.C., Arifuzzaman, M., Ranasinghe, M., Bhattacharya, S., Anker, J.N., Wang, G., and Li, C. (2021). Focused x-ray luminescence imaging system for small animals based on a rotary gantry. J. Biomed. Opt., 26. 5. Regularized reconstruction based on joint L 1 and total variation for sparse-view cone-beam X-ray luminescence computed tomography;Liu;Biomed. Opt. Express,2019
|
|