Biomechanical Behavior of a Variable Angle Locked Tibiotalocalcaneal Construct

Author:

Hamandi FarahORCID,Simon Gerard,Laughlin Richard,Goswami Tarun

Abstract

This paper examines the mechanics of the tibiotalocalcaneal construct made with a PHILOS plating system. A failed device consisting of the LCP plate and cortical, locking, and cannulated screws was used to perform the analysis. Visual, microstructure, and fractographic examinations were carried out to characterize the fracture surface topology. These examinations revealed the presence of surface scratching, inclusions, discoloration, corrosion pits, beach marks, and cleavage and striations on the fracture surface. Further examination of the material crystallography and texture revealed an interaction of S, Ni, and Mo-based inclusions that may have raised pitting susceptibility of the device made with Stainless Steel 316L. These features suggest that the device underwent damage by pitting the corrosion-fatigue mechanism and overloading towards the end to fail the plate and screws in two or more components. The screws failed via conjoint bending and torsion fatigue mechanisms. Computer simulations of variable angle locking screws were performed in this paper. The material of construction of the device was governed by ASTM F138-8 or its ISO equivalent 5832 and exhibited inconsistencies in chemistry and hardness requirements. The failure conditions were matched in finite element modeling and those boundary conditions discussed in this paper.

Publisher

MDPI AG

Subject

Bioengineering

Reference30 articles.

1. Surgical approach to bone healing in osteoporosis;Pe Pesce;Clin. Cases Miner. Bone Metab.,2009

2. Bone health and osteoporosis: A report of the surgeon general;McGowen;US Dep. Health Hum. Serv.,2004

3. Computational modelling of long bone fractures fixed with locking plates – How can the risk of implant failure be reduced?

4. Relative stability of conventional and locked plating fixation in a model of the osteoporotic femoral diaphysis

5. Prospects of implant with locking plate in fixation of subtrochanteric fracture: Experimental demonstration of its potential benefits on synthetic femur model with supportive hierarchical nonlinear hyperelastic finite element analysis;Latifi;Biomed. Eng.,2012

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献