Assessment of the Efficiency of Measuring Foot and Ankle Edema with a 3D Portable Scanner

Author:

Beldame Julien12,Sacco Riccardo3ORCID,Munoz Marie-Aude4,Masse Marion5,Lalevée Matthieu36

Affiliation:

1. ICP-Clinique Blomet, 136 bis Rue Blomet, 75015 Paris, France

2. Clinique Megival, 1328 Avenue de la Maison Blanche, 76550 Saint Aubin sur Scie, France

3. Department of Orthopedic Surgery, Rouen University Hospital, 37 Bd Gambetta, 76000 Rouen, France

4. Centre Médical Achille, 200 Avenue des Prés d’Arènes, 34070 Montpellier, France

5. CKS, Centre kiné sport Dieppe, 32 Rue Louis Blériot, Neuville les Dieppe, 76370 Dieppe, France

6. CETAPS EA3832, Research Center for Sports and Athletic Activities Transformations, University of Rouen Normandy, 76821 Mont-Saint-Aignan, France

Abstract

Background: To prospectively evaluate the reliability of a portable optical scanner compared to the water displacement technique for volumetric measurements of the foot and ankle and to compare the acquisition time associated with these two methods. Methods: Foot volume was measured in 29 healthy volunteers (58 feet, 24 females and 5 males) by a 3D scanner (UPOD-S 3D Laser Full-Foot Scanner®) and by water displacement volumetry. Measurements were performed on both feet, up to a height of 10 cm above the ground. The acquisition time for each method was evaluated. The Kolmogorov-Smirnov test, Lin’s Concordance Correlation Coefficient, and a Student’s t-test were performed. Results: Mean foot volume was 869.7 +/− 165.1 cm3 (3D scanner) versus 867.9 +/− 155.4 cm3 (water-displacement volumetry) (p < 10−5). The concordance of measurements was 0.93, indicative of a high correlation between the two techniques. Volumes were 47.8 cm3 lower when using the 3D scanner versus water volumetry. After statistically correcting this underestimation, the concordance was improved (0.98, residual bias = −0.03 +/− 35.1 cm3). The mean examination time was 4.2 +/− 1.7 min (3D optical scanner) versus 11.1 +/− 2.9 min (water volumeter) (p < 10−4). Conclusions: Ankle/foot volumetric measurements performed using this portable 3D scanner are reliable and fast and can be used in clinical practice and research.

Publisher

MDPI AG

Subject

Bioengineering

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3