In Vitro Wear of a Novel Vitamin E Crosslinked Polyethylene Lumbar Total Joint Replacement

Author:

Siskey Ryan L.1,Yarbrough Ronald V.2,Spece Hannah3,Hodges Scott D.2,Humphreys Steven C.2,Kurtz Steven M.3

Affiliation:

1. Exponent Inc., Philadelphia, PA 19104, USA

2. 3Spine, Chattanooga, TN 37402, USA

3. School of Biomedical Engineering, Science, and Health Systems, Drexel University, Philadelphia, PA 19104, USA

Abstract

Background: A novel, lumbar total joint replacement (TJR) design has been developed to treat degeneration across all three columns of the lumbar spine (anterior, middle, and posterior columns). Thus far, there has been no in vitro studies that establish the preclinical safety profile of the vitamin E-stabilized highly crosslinked polyethylene (VE-HXLPE) lumbar TJR relative to historical lumbar anterior disc replacement for the known risks of wear and impingement faced by all motion preserving designs for the lumbar spine. Questions/Purpose: In this study we asked, (1) what is the wear performance of the VE-HXLPE lumbar TJR under ideal, clean conditions? (2) Is the wear performance of VE-HXLPE in lumbar TJR sensitive to more aggressive, abrasive conditions? (3) How does the VE-HXLPE lumbar TJR perform under impingement conditions? Method: A lumbar TJR with bilateral VE-HXLPE superior bearings and CoCr inferior bearings was evaluated under clean, impingement, and abrasive conditions. Clean and abrasive testing were guided by ISO 18192-1 and impingement was assessed as per ASTM F3295. For abrasive testing, CoCr components were scratched to simulate in vivo abrasion. The devices were tested for 10 million cycles (MC) under clean conditions, 5 MC under abrasion, and 1 MC under impingement. Result: Wear rates under clean and abrasive conditions were 1.2 ± 0.5 and 1.1 ± 0.6 mg/MC, respectively. The VE-HXLPE components demonstrated evidence of burnishing and multidirectional microscratching consistent with microabrasive conditions with the cobalt chromium spherical counterfaces. Under impingement, the wear rates ranged between 1.7 ± 1.1 (smallest size) and 3.9 ± 1.1 mg/MC (largest size). No functional or mechanical failure was observed across any of the wear modes. Conclusions: Overall, we found that that a VE-HXLPE-on-CoCr lumbar total joint replacement design met or exceeded the benchmarks established by traditional anterior disc replacements, with wear rates previously reported in the literature ranging between 1 and 15 mg/MC. Clinical Relevance: The potential clinical benefits of this novel TJR design, which avoids long-term facet complications through facet removal with a posterior approach, were found to be balanced by the in vitro tribological performance of the VE-HXLPE bearings. Our encouraging in vitro findings have supported initiating an FDA-regulated clinical trial for the design which is currently under way.

Funder

3Spine

Publisher

MDPI AG

Subject

Bioengineering

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3