The Effect of Cooling Fluid Composition on Ablation Size in Hepatic Laser Ablation: A Comparative Study in an Ex Vivo Bovine Setting

Author:

Mankertz Fiona1ORCID,Bayerl Nadine2ORCID,Gemeinhardt Ole3ORCID,Hosten Norbert1ORCID,Kromrey Marie-Luise1ORCID

Affiliation:

1. Institute for Diagnostic Radiology and Neuroradiology, University Medicine Greifswald, 17475 Greifswald, Germany

2. Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität (FAU) Erlangen-Nürnberg, 91054 Erlangen, Germany

3. Department of Radiology, Charité—Universitätsmedizin Berlin, 10117 Berlin, Germany

Abstract

Purpose: Hyperthermic ablation is a minimally invasive mode of tumour therapy which serves as a viable alternative to surgical intervention. However, one of the major drawbacks, besides the heat sink effect and the risk of damaging adjacent organs, is limited ablation size. The use of a cooling fluid during ablation has been shown to increase the ablation volume and decrease the carbonisation rate. The aim of this study was to investigate whether the composition of the cooling fluid has an effect on ablation size and carbonisation rate during hepatic laser ablation in an ex vivo bovine setting. Method: In this study bovine hepatic tissue was ablated in an ex vivo setting using an internally cooled laser applicator. A total of 45 tissue samples were assigned to three groups: 0.9% saline infusion (n = 15), distilled water infusion (n = 15) and a 50%/50% mixture of 0.9% saline and distilled water (n = 15). Ablation was conducted using a 1064 nm Nd:YAG laser at a wattage of 25 W and time interval of 10 min. The ablation volume and carbonisation rate were then measured and recorded through postprocedural MRI. One-way ANOVA and post-hoc testing were performed to assess the effect of the cooling fluid composition on the ablation volumes. Results: We found that using a mixture of saline and distilled water as a cooling fluid during hyperthermic ablation resulted in a larger ablation volume (mean ± SD: 22.64 ± 0.99 cm3) when compared to saline infusion (21.08 ± 1.11 cm3) or distilled water infusion (20.92 ± 0.92 cm3). This difference was highly significant (p < 0.001). There was no significant difference in ablation size between the saline group and the distilled water group. The highest carbonisation rate occurred in the saline group (12/15), followed by the mixed infusion group (3/15) and the distilled water group (1/15). Conclusions: The results of this study suggest that cooling fluid composition during hepatic laser ablation affects ablation volume in an ex vivo bovine setting. There was no statistically significant difference when comparing ablation volumes during saline infusion and distilled water infusion, but the carbonisation rate was significantly higher when using saline. The combination of saline and distilled water in a 50%/50% mixture as cooling fluid appears to be an auspicious alternative, as ablation volumes created with it are larger when compared to saline and distilled water alone, while carbonisation rate remains low. This might improve patient outcome as well as patient eligibility for hyperthermic ablation.

Publisher

MDPI AG

Subject

Radiology, Nuclear Medicine and imaging

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3