Subcutaneous Implantable Cardioverter Defibrillator: A Contemporary Overview

Author:

Guarracini Fabrizio1ORCID,Preda Alberto2ORCID,Bonvicini Eleonora1,Coser Alessio1,Martin Marta1,Quintarelli Silvia1,Gigli Lorenzo2,Baroni Matteo2,Vargiu Sara2,Varrenti Marisa2ORCID,Forleo Giovanni Battista3ORCID,Mazzone Patrizio2ORCID,Bonmassari Roberto1ORCID,Marini Massimiliano1ORCID,Droghetti Andrea4ORCID

Affiliation:

1. Department of Cardiology, S. Chiara Hospital, 38122 Trento, Italy

2. Electrophysiology Unit, Cardio-Thoraco-Vascular Department, ASST Grande Ospedale Metropolitano Niguarda, 20162 Milan, Italy

3. Department of Thoracic Surgery, Candiolo Cancer Institute, FPO-IRCCS, Candiolo, 10060 Turin, Italy

4. Cardiology Unit, Luigi Sacco University Hospital, 20157 Milan, Italy

Abstract

The difference between subcutaneous implantable cardioverter defibrillators (S-ICDs) and transvenous ICDs (TV-ICDs) concerns a whole extra thoracic implantation, including a defibrillator coil and pulse generator, without endovascular components. The improved safety profile has allowed the S-ICD to be rapidly taken up, especially among younger patients. Reports of its role in different cardiac diseases at high risk of SCD such as hypertrophic and arrhythmic cardiomyopathies, as well as channelopathies, is increasing. S-ICDs show comparable efficacy, reliability, and safety outcomes compared to TV-ICD. However, some technical issues (i.e., the inability to perform anti-bradycardia pacing) strongly limit the employment of S-ICDs. Therefore, it still remains only an alternative to the traditional ICD thus far. This review aims to provide a contemporary overview of the role of S-ICDs compared to TV-ICDs in clinical practice, including technical aspects regarding device manufacture and implantation techniques. Newer outlooks and future perspectives of S-ICDs are also brought up to date.

Publisher

MDPI AG

Subject

Paleontology,Space and Planetary Science,General Biochemistry, Genetics and Molecular Biology,Ecology, Evolution, Behavior and Systematics

Reference98 articles.

1. Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients;Alter;Pacing Clin. Electrophysiol.,2005

2. An Entirely Subcutaneous Implantable Cardioverter–Defibrillator;Bardy;N. Engl. J. Med.,2010

3. Clinical experience with a novel subcutaneous implantable defibrillator system in a single center;Theuns;Clin. Res. Cardiol.,2011

4. Procedure, management, and outcome of subcutaneous implantable cardioverter–defibrillator extraction in clinical practice;Migliore;EP Eur.,2023

5. Comparison of transvenous vs subcutaneous defibrillator therapy in patients with cardiac arrhythmia syndromes and genetic cardiomyopathies;Kuschyk;Int. J. Cardiol.,2021