Abstract
We evaluated the feasibility of myocardial perfusion ratio to the aorta (MPR) in static computed tomography perfusion (CTP) for detecting myocardial perfusion abnormalities assessed by single-photon emission computed tomography (SPECT). Twenty-five patients with suspected coronary artery disease who underwent dynamic CTP and SPECT were retrospectively evaluated. CTP images scanned at a sub-optimal phase for detecting myocardial perfusion abnormalities were selected from dynamic CTP images and used as static CTP images in the present study. The diagnostic accuracy of MPR derived from static CTP was compared to those of visual assessment and conventional quantitative parameters such as myocardial CT attenuation (HU) and transmural perfusion ratio (TPR). The area under the curve of MPR (0.84; 95% confidence interval [CI], 0.76–0.90) was significantly higher than those of myocardial CT attenuation (0.73; 95% CI, 0.65–0.79) and TPR (0.76; 95% CI, 0.67–0.83) (p < 0.05). Sensitivity and specificity were 67% (95% CI, 54–77%) and 90% (95% CI, 86–92%) for visual assessment, 51% (95% CI, 39–63%) and 86% (95% CI, 82–89%) for myocardial CT attenuation, 63% (95% CI, 51–74%) and 84% (95% CI, 80–88%) for TPR, and 78% (95% CI, 66–86%) and 84% (95% CI, 80–88%) for MPR, respectively. MPR showed higher diagnostic accuracy for detecting myocardial perfusion abnormality compared with myocardial CT attenuation and TPR.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献