Abstract
The aim was to evaluate the use of a bovine procalcitonin (PCT) ELISA kit (Cusabio, China) for assessing PCT in bovine milk samples. Validation was performed by using 10 plasma and corresponding milk samples from mastitic cows. The limit of detection (LOD) was calculated. The coefficient of variation (CV%) of the readings of five plasma samples measured five times in the same plate (intra-assay) and the CV% of the same five samples read five times in three separate plates was evaluated. Parallelism was determined by serial twofold dilutions of five plasma and corresponding milk samples. Milk samples were analyzed with and without centrifugation. Regarding plasma PCT, the method presented an inter- and intra-CV < 23.7% and parallelism had very good recovery values. The ELISA kit studied can measure bovine plasma PCT concentrations. The kit antibodies fail in binding PCT in milk samples because all centrifuged milk samples showed a lower LOD than blank samples. Only three uncentrifuged milk samples showed measurable PCT concentrations. Due to these results, the commercial ELISA kit investigated could not be employed for the detection of PCT in milk samples.
Subject
General Veterinary,Animal Science and Zoology
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献