Performance of Austenitic High-Nitrogen Steels under Gross Slip Fretting Corrosion in Bovine Serum

Author:

Fischer Alfons12ORCID,Telouk Philipe3ORCID,Beckmann Christian4ORCID,Heermant Saskia4,Wittrock Adrian4ORCID,Debus Jörg4ORCID,Wimmer Markus A.2ORCID

Affiliation:

1. Max Planck Institute for Sustainable Materials, Microstructure Physics and Alloy Design, 40237 Duesseldorf, Germany

2. Department of Orthopedic Surgery, Rush University Medical Center, Chicago, IL 60612, USA

3. Laboratoire de Géologie, Université de Lyon, 69342 Lyon, France

4. Department of Physics, TU Dortmund University, 44227 Dortmund, Germany

Abstract

Modular artificial hip joints are a clinical standard today. However, the release of wear products from the head–taper interface, which includes wear particles in the nm size range, as well as metal ions, have raised concerns. Depending on the loading of such taper joints, a wide variety of different mechanisms have been found by retrieval analyses. From these, this paper concentrates on analyzing the contribution of gross slip fretting corrosion at ultra-mild wear rates using a bovine calf serum solution (BCS) as the lubricant. The parameters were chosen based on biomechanical considerations, producing wear rates of some ng/m wear path. In parallel, the evolution of tribomaterial (third bodies) was analyzed as to its constituents and generation rates. It has already been shown earlier that, by an advantageous combination of wear mechanisms and submechanisms, certain constituents of the tribomaterial remain inside the contact area and act like extreme-pressure lubricant additives. For the known wear and corrosion resistance of austenitic high-nitrogen steels (AHNSs), which outperform CoCrMo alloys even under inflammatory conditions, we hypothesized that such steels will generate ultra-mild wear rates under gross slip fretting. While testing AHNSs against commercially available biomedical-grade materials of CoCrMo and TiAlV alloys, as well as zirconia-toughened alumina (ZTA) and against itself, it was found that AHNSs in combination with a Ti6Al4V alloy generated the smallest wear rate under gross slip fretting corrosion. This paper then discusses the wear behavior on the basis of ex situ analyses of the worn surfaces as to the acting wear mechanisms and submechanisms, as well as to the tribological reaction products.

Funder

Galante Family Foundation, Niles, IL, US

Grainger Chair of the Arthritis and Orthopedics Institute, Chicago, IL, US

AlCove Management GmbH, Essen, Germany

Publisher

MDPI AG