Abstract
Heart rate (HR) and heart rate variability (HRV) based physiological metrics such as Excess Post-exercise Oxygen Consumption (EPOC), Energy Expenditure (EE), and Training Impulse (TRIMP) are widely utilized in coaching to monitor and optimize an athlete’s training load. Chest straps, and recently also dry electrodes integrated to special sports vests, are used to monitor HR during sports. Mechanical design, placement of electrodes, and ergonomics of the sensor affect the measured signal quality and artefacts. To evaluate the impact of the sensor mechanical design on the accuracy of the HR/HRV and further on to estimation of EPOC, EE, and TRIMP, we recorded HR and HRV from a chest strap and a vest with the same ECG sensor during supervised exercise protocol. A 3-lead clinical Holter ECG was used as a reference. Twenty-five healthy subjects (six females) participated. Mean absolute percentage error (MAPE) for HR was 0.76% with chest strap and 3.32% with vest. MAPE was 1.70% vs. 6.73% for EE, 0.38% vs. 8.99% for TRIMP and 3.90% vs. 54.15% for EPOC with chest strap and vest, respectively. Results suggest superior accuracy of chest strap over vest for HR and physiological metrics monitoring during sports.