Abstract
We present a noise estimation and subtraction algorithm capable of increasing the sensitivity of heterodyne laser interferometers by one order of magnitude. The heterodyne interferometer is specially designed for dynamic measurements of a test mass in the application of sub-Hz inertial sensing. A noise floor of 3.31×10−11m/Hz at 100 mHz is achieved after applying our noise subtraction algorithm to a benchtop prototype interferometer that showed a noise level of 2.76×10−10m/Hz at 100 mHz when tested in vacuum at levels of 3×10−5 Torr. Based on the previous results, we investigated noise estimation and subtraction techniques of non-linear optical pathlength noise, laser frequency noise, and temperature fluctuations in heterodyne laser interferometers. For each noise source, we identified its contribution and removed it from the measurement by linear fitting or a spectral analysis algorithm. The noise correction algorithm we present in this article can be generally applied to heterodyne laser interferometers.
Funder
National Geospatial-Intelligence Agency
National Science Foundation
Jet Propulsion Laboratory
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Reference30 articles.
1. The LTP interferometer and phasemeter
2. LISA Phasemeter development;Wand,2006
3. An Interferometric Based Optical Read-Out Scheme For The LISA Proof-Mass;Aston,2006
4. LISA Pathfinder: the experiment and the route to LISA
Cited by
20 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献