Characterization of Micro-Threaded Stem Taper Surfaces of Cementless Hip Endoprostheses

Author:

Dolinar Drago12,Kocjančič Boštjan12,Avsec Klemen12,Šetina Batič Barbara3ORCID,Kocijan Aleksandra3ORCID,Godec Matjaž3ORCID,Sedlaček Marko3ORCID,Debeljak Mojca4ORCID,Grant John T.5ORCID,Zupanc Timon1,Jenko Monika67

Affiliation:

1. Department for Orthopaedic Surgery, University Medical Centre Ljubljana, Zaloška 9, 1000 Ljubljana, Slovenia

2. Orthopaedic Surgery Chair, Faculty of Medicine, University of Ljubljana, Vrazov trg 2, 1000 Ljubljana, Slovenia

3. Institute of Metals and Technology, Lepi pot 11, 1000 Ljubljana, Slovenia

4. University Rehabilitation Institute Republic of Slovenia, Linhartova 51, 1000 Ljubljana, Slovenia

5. Research Institute, University of Dayton, Dayton, OH 45469, USA

6. MD-RI Institute for Materials Research in Medicine, Bohoričeva 5, 1000 Ljubljana, Slovenia

7. MD Medicina, Sanatorium, Bohoričeva 5a, 1000 Ljubljana, Slovenia

Abstract

We investigated micro-threaded stem taper surface and its impact on premature failures, aseptic loosening, and infection in cementless hip endoprostheses. Our study focused on the fretting, and crevice corrosion of micro-threaded tapers, as well as the characterization of the microstructure and surface properties of two new and three retrieved Zweymüller stem tapers. The retrieved samples were selected and examined based on the head–stem taper interface being the sole source of modularity with a metallic component, specifically between the Ti alloy taper stem and the ceramic head. To determine the surface chemistry and microstructures of both new and retrieved hip endoprostheses stem taper titanium alloy, scanning -electron microscopy (SEM) was employed for morphological and microstructural analyses. Energy dispersive spectroscopy (EDS) was utilized for characterizing chemical element distribution, and electron backscattered diffraction (EBSD) was used for phase analysis. The roughness of the micro-threated stem tapers from different manufacturers was investigated using an optical profilometer, with standard roughness parameters Ra (average surface roughness) and Rz (mean peak to valley height of the roughness profile) being measured. Electrochemical studies revealed no fretting corrosion in retrieved stem tapers with ceramic heads. Consequently, three retrieved tapers and two new ones for comparison underwent potentiodynamic measurements in Hank’s solution to determine the corrosion rate of new and retrieved stem taper surfaces. The results showed a low corrosion rate for both new and prematurely failed retrieved samples due to aseptic loosening. However, the corrosion rate was higher in infected and low-grade infected tapers. In conclusion, our study suggests that using ceramic heads reduces taper corrosion and subsequently decreases the incidence of premature failures in total hip arthroplasty.

Funder

Slovenian Research Agency, ARIS

Department for Orthopedic Surgery UMC, Ljubljana, Slovenia, Tertiary projects