Free Myocutaneous Flap Assessment in a Rat Model: Verification of a Wireless Bioelectrical Impedance Assessment (BIA) System for Vascular Compromise Following Microsurgery

Author:

Huang Yao-KuangORCID,Wong Min Yi,Wu Chi-Rung,Cheng Yung-Ze,Lin Bor-ShyhORCID

Abstract

Background: Microvascular tissue transfer is a common reconstructive procedure. We designed a bioelectrical impedance assessment (BIA) system for quantitative analysis of tissue status. This study attempts to verify it through the animal model. Methods: The flaps of the rat model were monitored by the BIA system. Results: The BIA variation of the free flap in the rat after the vascular compromise was recorded. The non-vascular ligation limbs of the same rat served as a control group. The bio-impedance in the experimental group was larger than the control group. The bio-impedances of both the thigh/feet flaps in the experimental group were increased over time. In the thigh, the difference in bio-impedance from the control group was first detected at 10 kHz at the 3rd and last at 1 kHz at the 6th h, after vascular compromise. The same finding was observed in the feet. Compared with the control group, the bio-impedance ratio (1 kHz/20 kHz) of the experimental group decreased with time, while their variation tendencies in the thigh and feet were similar. Conclusions: The flap may be monitored by the BIA for vascular status.

Funder

Ministry of Science and Technology in Taiwan

Publisher

MDPI AG

Subject

Medicine (miscellaneous)

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3