Abstract
(1) Background: The purpose of this study is to evaluate the impact of an augmented reality navigation system (SIRIO) for percutaneous biopsies and ablative treatments on bone lesions, compared to a standard CT-guided technique. (2) Methods: Bioptic and ablative procedures on bone lesions were retrospectively analyzed. All procedures were divided into SIRIO and Non-SIRIO groups and in <2 cm and >2 cm groups. Number of CT-scans, procedural time and patient’s radiation dose were reported for each group. Diagnostic accuracy was obtained for bioptic procedures. (3) Results: One-hundred-ninety-three procedures were evaluated: 142 biopsies and 51 ablations. Seventy-four biopsy procedures were performed using SIRIO and 68 under standard CT-guidance; 27 ablative procedures were performed using SIRIO and 24 under standard CT-guidance. A statistically significant reduction in the number of CT-scans, procedural time and radiation dose was observed for percutaneous procedures performed using SIRIO, in both <2 cm and >2 cm groups. The greatest difference in all variables examined was found for procedures performed on lesions <2 cm. Higher diagnostic accuracy was found for all SIRIO-assisted biopsies. No major or minor complications occurred in any procedures. (4) Conclusions: The use of SIRIO significantly reduces the number of CT-scans, procedural time and patient’s radiation dose in CT-guided percutaneous bone procedures, particularly for lesions <2 cm. An improvement in diagnostic accuracy was also achieved in SIRIO-assisted biopsies.
Cited by
11 articles.
订阅此论文施引文献