Abstract
We compared the results of absorbed dose measurements made using the TRS-398, TG-51, and DIN protocols and their associated uncertainties to reduce discrepancies in measurement results made using the three protocols. This experiment was carried out on two Varian Medical linear accelerators with 4, 6, 10, and 20 MV photon energies using FC65-G and CC15 (cylindrical) and NACP-02-type (plane-parallel) ion chambers in water phantoms. The radiation beam quality index (Q) was determined from the measurement of percentage depth dose. It was used to determine the photon beam quality factor required with the ionization chamber calibration factor to convert the ion chamber reading into the absorbed dose to water. For the same beam quality, the TRS-398/TG-51 varied from 0.01% to 1.8%, whereas the ratio for TRS-398/DIN 6800-2 varied from 0.1% to 0.88%. The chamber-to-chamber variation was 0.09% in TRS-398/TG-51, 0.03% in TRS-398/DIN, and 0.02% in TG-51/DIN 6800-2. The expanded uncertainties (k = 1) were 1.24 and 1.25 when using TRS-398 and DIN 6800-2, respectively. Using the aforementioned three protocols, the results showed little chamber-to-chamber variation and uncertainty in absorbed dose measurements. The estimated uncertainties when using cylindrical ion chambers were slightly lower than those measured using plane-parallel chambers. The results are important in facilitating comparisons of absorbed dose measurements when using the three protocols.
Funder
Imam Mohammad Ibn Saud Islamic University
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Reference20 articles.
1. AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
2. Dosismessverfahren Nach der Sondenmethode für Photonen—Und Elektronenstrahlung Teil 2: Ionisationsdosimetrie DIN 6800-2,1998
3. Code of Practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration service
4. Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams,2008
5. Absorbed dose determination in external beam radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water;Andreo,2000
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献