Abstract
We present the results of ex vivo exposure by an ultrafast all-fiber Holmium laser system to porcine longissimus muscle tissues. A simple Ho-doped laser system generated ultrashort pulsed radiation with less than 1 ps pulse width and a repetition rate of 20 MHz at a central wavelength of 2.06 μm. Single-spot ex vivo experiments were performed at an average power of 0.3 W and different exposure times of 5, 30 and 60 s, varying the total applied energy in the range of 1.5–18 J. Evaluation of laser radiation exposure was performed according to the depth and diameter of coagulation zones, ablation craters and thermal damage zones during the morphological study. Exposure by ultrashort pulsed radiation with an average power of 0.3 W showed destructive changes in the muscle tissue after 5 s and nucleation of an ablative crater. The maximum ablation efficiency was about 28% at the ablation depth and diameter of 180 μm and 500 μm, respectively. The continuous-wave radiation impact at the same parameters resulted only in heating of the near-muscular tissue, without ablation and coagulation traces. Exposure to tissue with an average power at 0.3 W of ultrashort pulsed radiation led, within 30 and 60 s, to similar results as caused by 0.5 W of continuous-wave radiation, although with less carbonization formation.
Funder
The Ministry of Science and Higher Education of Russian Federation
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献