Comparative Study on the Responsiveness of Different Species to Coagulation and Complement for Hemocompatibility Evaluation

Author:

Kim Jeonghwa1ORCID,Kim Geonyong1,Lim Jiyoung2,Kim Sekyung2,Eom Joonho2,Kim Taewon1

Affiliation:

1. Bio-Health Center (GLP), Korea Testing Certification Institute, Cheongju 28115, Republic of Korea

2. Department of Medical Device Research Division, National Institute of Food and Drug Safety Evaluation, Cheongju 28159, Republic of Korea

Abstract

Background: According to the International Standard (ISO10993-4), a test model using human blood must be used for blood compatibility evaluation of medical devices. However, various safety investigations have reported outcomes from animal models simulating clinical conditions. Additionally, the validation of animal blood in blood compatibility assessment models is limited. In this study, the differences in the reactivity of different species to coagulation and complement, as well as the main mechanisms governing blood clot formation upon contact with medical devices, were examined. Moreover, it aimed to acquire information required to design a hemocompatibility evaluation model and interpret the results. Results: Rabbits, porcine, rhesus monkeys, and cynomolgus monkeys were exposed to one negative and two positive control substances, and variations in the partial thromboplastin time (PTT) were observed. The activity of the complement system was observed in accordance with the test method in ISO 10993-4. Although the PTT absolute values varied across animal species, the normalized PTT values—61% for rabbits, 66% for pigs, 63% for rhesus monkeys, and 65% for cynomolgus monkeys—exhibited no statistically significant differences. After reacting human and porcine sera with the material, the test model using human serum distinguished the positive and negative control candidates, whereas the test model using porcine serum could not discriminate them. Conclusions: These results suggest similar reactivity in blood coagulation across species after substance contact. However, complement system activity indicated a significant difference in reactivity between human and porcine blood. This finding will be useful in the design of a blood compatibility evaluation model for medical devices and interpretation of the results.

Funder

Ministry of Food and Drug Safety

Publisher

MDPI AG

Reference32 articles.

1. (2017). Biological Evaluation of Medical Devices—Part 4: Selection of Tests for Interactions with Blood (Standard No. ISO 10993-4).

2. (2019). Standard Practice for Platelet Leukocyte Count-An In-Vitro Measure for Hemocompatibility Assessment of Cardiovascular Materials (Standard No. ASTM F2888-19).

3. (2016). Cardiovascular Implants and Extracorporeal Systems—Vascular Prostheses—Tubular Vascular Grafts and Vascular Patches (Standard No. ISO 7198:2016).

4. (2016). Cardiovascular Implants and Artificial Organs—Blood-Gas Exchangers (Oxygenators) (Standard No. ISO 7199:2016).

5. Some species differences in fibrinolysis and blood coagulation;Mason;J. Biomed. Mater. Res.,1971

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3