This study aimed to determine the validity of specific knee varus alignment measurement methods. We measured the femorotibial angle (FTA) using radiography and optical motion capture and validated the FTA measurement using markerless motion capture. The subjects included 34 legs of 19 patients with knee osteoarthritis (OA). One-way analysis of variance and multiple comparison tests were used to compare the FTA values between the Kellgren–Lawrence classification (KL) and Pearson’s correlation coefficient for validity. The analysis showed that the FTA measured by markerless motion capture had a significant correlation to the FTA measured by radiography (r = 0.869, p < 0.01) and significantly increased with increasing KL (p < 0.05). These results indicate that markerless motion capture is a valid outcome measure for varus alignment in patients with knee OA.

Publisher

MDPI AG

Subject

Geriatrics and Gerontology,Gerontology,Aging,Health (social science)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3