Abstract
Prospective studies applying fluoroscopy for assessment of right ventricular (RV) lead position have failed to show clear benefits from RV septal pacing. We investigated the impact of different RV lead positions verified by computed tomography (CT) on the risk of pacing-induced cardiomyopathy (PICM). We retrospectively included 153 patients who underwent routine fluoroscopy-guided pacemaker implantation between March 2012 and May 2020. All patients had normal pre-implant left ventricular ejection fraction (LVEF). Patients attended a follow-up visit including contrast-enhanced cardiac CT and transthoracic echocardiography. Patients were classified as septal or non-septal based on CT analysis. The primary endpoint was PICM (LVEF < 50% with ≥10% decrease after implantation). Based on CT, 48 (31.4%) leads were septal and 105 (68.6%) were non-septal. Over a median follow-up of 3.1 years, 16 patients (33.3%) in the septal group developed PICM compared to 31 (29.5%) in the non-septal group (p = 0.6). Overall, 13.1% deteriorated to LVEF ≤ 40%, 5.9% were upgraded to cardiac resynchronization therapy device, and 14.4% developed new-onset atrial fibrillation, with no significant differences between the groups. This study demonstrated a high risk of PICM despite normal pre-implant left ventricular systolic function with no significant difference between CT-verified RV septal or non-septal lead position.
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献