Abstract
Fall rates are increasing among the aging population and even higher falls rates have been reported in populations with neurological impairments. The Berg Balance Scale is often used to assess balance in older adults and has been validated for use in people with stroke, traumatic brain injury, and Parkinson’s disease. While the Berg Balance Scale (BBS) has been found to be predictive of the length of rehabilitation stay following stroke, a recent review concluded the BBS lacked predictive validity for fall risk. Conversely, sophisticated measures assessing center of mass (COM) displacement have shown to be predictive of falls risk. However, calculating COM displacement is difficult to measure outside a laboratory. Accordingly, we sought to validate COM displacement measurements derived from an HTC Vive tracker secured to the pelvis by comparing it to COM derived from ‘gold’ standard laboratory-based full-body motion capture. Results showed that RMS between the COM calculated from HTC Vive tracker and full body motion capture agree with an average error rate of 2.1 ± 2.6 cm. Therefore, we conclude measurement of COM displacement using an HTC Vive tracker placed on the pelvis is reasonably representative of laboratory-based measurement of COM displacement.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献