Abstract
We describe a control system for operating valve-enabled microfluidic devices and leverage this control system to carry out a complex workflow of plasma separation from 8 μL of whole blood followed by on-chip mixing of plasma with assay reagents for biomarker detection. The control system incorporates pumps, digital pressure sensors, a microcontroller, solenoid valves and off-the-shelf components to deliver high and low air pressure in the desired temporal sequence to meter fluid flow and actuate microvalves. Importantly, our control system is portable, which is suitable for operating the microvalve-enabled microfluidic devices in the point-of-care setting.
Funder
NIH
Center for Biomedical Discovery, the Office of Translation to Practice and Cells to Cures Strategic Initiative at Mayo Clinic
Subject
Clinical Biochemistry,General Medicine,Analytical Chemistry,Biotechnology,Instrumentation,Biomedical Engineering,Engineering (miscellaneous)
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献