Fatigue Design of Dental Implant Assemblies: A Nominal Stress Approach

Author:

Armentia Mikel,Abasolo Mikel,Coria Ibai,Albizuri JosebaORCID

Abstract

Fatigue is the most common mechanical failure type in dental implants. ISO 14801 standardizes fatigue testing of dental implants, providing the load-life curve which is most useful for comparing the fatigue behavior of different dental implant designs. Based on it, many works were published in the dental implant literature, comparing different materials, component geometries, connection types, surface treatments, etc. These works are useful for clinicians in order to identify the best options available in the market. The present work is intended not for clinicians but for dental implant manufacturers, developing a design tool that combines Finite Element Analysis, fatigue formulation and ISO 14801 experimental tests. For that purpose, 46 experimental tests were performed on BTI INTERNA® IIPSCA4513 implants joined with INPPTU44 abutments by means of INTTUH prosthetic screws under three different tightening torque magnitudes. Then, the load case was reproduced in a FE model from where the nominal stress state in the fatigue critical section was worked out. Finally, Walker criterion was used to represent accurately the effects of mean stress and predict fatigue life of the studied dental implant assembly, which can be extended to most of the products of BTI manufacturer. By means of this tool, dental implant manufacturers will be able to identify the critical design and assembly parameters in terms of fatigue behavior, evaluate their influence in preliminary design stages and consequently design dental implants with significantly better fatigue response which in turn will reduce future clinical incidences.

Funder

Eusko Jaurlaritza

Publisher

MDPI AG

Subject

General Materials Science,Metals and Alloys

Reference44 articles.

1. An introduction to single implant abutments

2. Theoretical analysis of the fatigue life of fixture screws in osseointegrated dental implants;Patterson;Int. J. Oral Maxillofac. Implant.,1992

3. Introduction to the Design and Behavior of Bolted Joints;Bickford,2008

4. Preload and torque removal evaluation of three different abutment screws for single standing implant restorations

5. Finite element analysis to determine implant preload