Affiliation:
1. Department of Radiology, New York University Grossman School of Medicine, 660 First Avenue, Room 240, New York, NY 10016, USA
Abstract
A novel gas chromatography method was developed using automatic injections to identify and quantify the amount of residual solvents or analytes in samples of fluorine-18 and carbon-11 radiopharmaceuticals. This approach evaluates seven analytes in less than 5 versus 13 min of acquisition time. The method additionally includes a 3 min bakeout to aid in the removal and carry-over of higher-boiling impurities. Chromatographic parameters such as column temperature, hold time, column pressure, flow rate, and split ratios were adjusted and optimized to analyze radioactive drug samples containing analytes which include methanol, ethanol, acetone, acetonitrile, triethylamine, N,N-dimethylformamide, and dimethyl sulfoxide. The relative standard deviation for each solvent was determined to be no greater than 1.6%. The method limit of detection (LOD) and limit of quantification (LOQ) were between 0.053 and 0.163 and 0.000 (5.791 × 10−6) and 0.520 mg/mL, respectively. This GC technique, using flame ionization detection (FID), was validated and is currently employed for the routine quality control of all approved IND and RDRC PET radiopharmaceuticals at our center.
Subject
Drug Discovery,Pharmaceutical Science,Molecular Medicine