Abstract
Accurate analysis using a simple and rapid procedure is always the most important pursuit of analytical chemists. In this study, a new sample preparation procedure, namely the shaker cup (SH) method, was designed and compared with two sample preparation procedures, commonly used in the laboratory, from three aspects: homogeneity of the sample–flux mixture, potential for sample contamination, and sample preparation time. For the three methods, a set of 54 certified reference materials (CRMs) was used to establish the calibration curves, while another set of 19 CRMs was measured to validate the results. In the calibration procedures, the matrix effects were corrected using the theoretical alpha coefficient method combined with the experimental coefficient method. The data of the major oxides (SiO2, TiO2, Al2O3, TFe2O3, MnO, MgO, CaO, Na2O, K2O, and P2O5) and minor elements (Cr, Cu, Ba, Ni, Sr, V, Zr, and Zn) obtained by wavelength dispersive X-ray fluorescence spectroscopy (WD-XRF) were compared using two derivative equations based on the findings by Laurence Whitty-Léveillé. The results revealed that the WD-XRF measured values using the SH method best agreed with the values recommended in the literature.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Reference55 articles.
1. Application and analysis of portable X-ray fluorescence analyzer on fluorite exploration in shallow cover area;Li;Comput. Tech. Geophys. Geochem. Explor.,2018
2. Application progress of X-ray fluorescence spectroscopy in the detection of heavy metals in soil;Meng;Food Mach.,2017
3. Capabilities of elemental analysis by EDXRF for geochemistry
4. Determination of S, Cl and F in silicate rocks by X-ray fluorescence analyses
5. Determination of trace chromium in cement by X-ray fluorescence spectrometer;Ren;Chin. J. Anal. Lab.,2015
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献