Author:
Berard David,Sen Chirantan,Nawn Corinne D.,Blackburn August N.,Ryan Kathy L.,Blackburn Megan B.
Abstract
Endotracheal intubation is a common life-saving procedure implemented in emergency care to ensure patient oxygenation, but it is difficult and often performed in suboptimal conditions leading to high rates of patient complications. Undetected misplacement in the esophagus is a preventable complication that can lead to fatalities in 5–10% of patients who undergo emergency intubation. End-tidal carbon dioxide monitoring and other proper placement detection methods are useful, yet the problem of misplacement persists. Our previous work demonstrated the utility of spectral reflectance sensors for differentiating esophageal and tracheal tissues, which can be used to confirm proper endotracheal tube placement. In this study, we examine the effectiveness of spectral characterization in the presence of saline, blood, “vomit”, and soot in the trachea. Our results show that spectral properties of the trachea that differentiate it from the esophagus persist in the presence of these substances. This work further confirms the potential usefulness of this novel detection technology in field applications.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
2 articles.
订阅此论文施引文献