The surface electromyography (SEMG) is one of the most popular bio-signals that can be applied in health monitoring systems, fitness training, and rehabilitation devices. Commercial clothing embedded with textile electrodes has already been released onto the market, but there is insufficient information on the performance of textile SEMG electrodes because the required configuration may differ according to the electrode material. The current study analyzed the influence of electrode size and pattern reduction rate (PRR), and hence the clothing pressure (Pc) based on in vivo SEMG signal acquisition. Bipolar SEMG electrodes were made in different electrode diameters Ø 5–30 mm, and the clothing pressure ranged from 6.1 to 12.6 mmHg. The results supported the larger electrodes, and Pc showed better SEMG signal quality by showing lower baseline noise and a gradual increase in the signal to noise ratio (SNR). In particular, electrodes, Ø ≥ 20 mm, and Pc ≥ 10 mmHg showed comparable performance to Ag-Ag/Cl electrodes in current textile-based electrodes. The current study emphasizes and discusses design factors that are particularly required in the designing and manufacturing process of smart clothing with SEMG electrodes, especially as an aspect of clothing design.