Abstract
Respiration monitoring is vital for human health assessment. Humidity sensing is a promising way to establish a relationship between human respiration and electrical signal. This paper presents a polyimide-based film bulk acoustic resonator (PI-FBAR) humidity sensor operating in resonant frequency and reflection coefficient S11 dual-parameter with high sensitivity and stability, and it is applied in real-time human respiration monitoring for the first time. Both these two parameters can be used to sense different breathing conditions, such as normal breathing and deep breathing, and breathing with different rates such as normal breathing, slow breathing, apnea, and fast breathing. Experimental results also indicate that the proposed humidity sensor has potential applications in predicting the fitness of individual and in the medical field for detecting body fluids loss and daily water intake warning. The respiratory rates measured by our proposed PI-FBAR humidity sensor operating in frequency mode and S11 mode have Pearson correlation of up to 0.975 and 0.982 with that measured by the clinical monitor, respectively. Bland–Altman method analysis results further revealed that both S11 and frequency response are in good agreement with clinical monitor. The proposed sensor combines the advantages of non-invasiveness, high sensitivity and high stability, and it has great potential in human health monitoring.
Funder
National Key Research and Development Project
National Natural Science Foundation of China
Subject
Electrical and Electronic Engineering,Mechanical Engineering,Control and Systems Engineering
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献