Abstract
Background: Leadless transcatheter pacemaker systems (TPS) have become a valuable alternative to transvenous pacemakers in selected indications. With the steadily increasing amount of TPS implantations performed worldwide, reports of periprocedural complications are likewise increasingly found in the literature but are still underreported. Case presentation: We report a case of a 75 year old male undergoing TPS implantation due to cardioinhibitory vasovagal syncope. The implantation was primarily uneventful; adequate pacing parameters and fixation of the device were achieved. Unfortunately, dislocation of the leadless pacemaker occurred at the end of the procedure and the device embolized into a primary side branch of the right pulmonary artery. Endovascular retrieval was performed by using a single snare technique without any further complications. Conclusions: Although challenging, endovascular recovery of embolized TPS from the pulmonary artery is feasible and may be successfully accomplished by experienced implanters.
Subject
Pharmacology (medical),General Pharmacology, Toxicology and Pharmaceutics
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献