A Comparison of Conical and Cylindrical Implants Inserted in an In Vitro Post-Extraction Model Using Low-Density Polyurethane Foam Blocks

Author:

Comuzzi Luca1,Tumedei Margherita23,Di Pietro Natalia45ORCID,Romasco Tea45ORCID,Heydari Sheikh Hossein Hamid456ORCID,Montesani Lorenzo7,Inchingolo Francesco8ORCID,Piattelli Adriano910,Covani Ugo11

Affiliation:

1. Independent Researcher, San Vendemiano-Conegliano, 31020 Treviso, Italy

2. Department of Medical, Surgical, and Dental Sciences, University of Milan, 20122 Milan, Italy

3. Maxillo-Facial Surgery and Dental Unit, Fondazione IRCCS Ca’ Granda Ospedale Maggiore Policlinico, 20122 Milan, Italy

4. Department of Medical, Oral and Biotechnological Sciences, “G. D’Annunzio” University of Chieti-Pescara, 66013 Chieti, Italy

5. Center for Advanced Studies and Technology-CAST, “G. D’Annunzio” University of Chieti-Pescara, 66013 Chieti, Italy

6. Villa Serena Foundation for Research, Via Leonardo Petruzzi 42, 65013 Città Sant’Angelo, Italy

7. Independent Researcher, 00187 Rome, Italy

8. Interdisciplinary Department of Medicine, University of Bari “Aldo Moro”, 70121 Bari, Italy

9. School of Dentistry, Saint Camillus International University of Health and Medical Sciences, 00131 Rome, Italy

10. Facultad de Medicina, UCAM Universidad Católica San Antonio de Murcia, 30107 Murcia, Spain

11. Department of Stomatology, Tuscan Stomatologic Institute, Foundation for Dental Clinic, Research and Continuing Education, 55041 Camaiore, Italy

Abstract

Combining tooth extraction and implant placement reduces the number of surgical procedures that a patient must undergo. Thus, the present study aimed to compare the stability of two types of conical implants (TAC and INTRALOCK) and another cylindrical one (CYROTH), inserted with a range of angulation of 15–20 degrees in low-density polyurethane blocks (10 and 20 pounds per cubic foot, PCF) with or without a cortical lamina (30 PCF), which potentially mimicked the post-extraction in vivo condition. For this purpose, a total of 120 polyurethane sites were prepared (10 for each implant and condition) and the Insertion Torque (IT), Removal Torque (RT), and Resonance Frequency Analysis (RFA) were measured, following a Three-Way analysis of variance followed by Tukey’s post hoc test for the statistical analysis of data. The IT and RT values registered for all implant types were directly proportional to the polyurethane density. The highest IT was registered by INTRALOCK implants in the highest-density block (32.44 ± 3.28 Ncm). In contrast, the highest RFA, a well-known index of Implant Stability Quotient (ISQ), was shown by TAC implants in all clinical situations (up to 63 ISQ in the 20 PCF block without the cortical sheet), especially in lower-density blocks. Although more pre-clinical and clinical studies are required, these results show a better primary stability of TAC conical implants in all tested densities of this post-extraction model, with a higher ISQ, despite their IT.

Publisher

MDPI AG

Subject

General Materials Science

Reference36 articles.

1. Recent Advancements in Nanomaterials for Biomedical Implants;Gautam;Biomed. Eng. Adv.,2022

2. Socket Healing with and without Immediate Implant Placement;Silva;Periodontology 2000,2019

3. Immediate Implant Placement: Current Concepts;Dhami;J. Nepal. Soc. Periodontol. Oral Implantol.,2019

4. The Tübinger Immediate Implant;Schulte;Quintessenz,1976

5. Implant Placement Post Extraction in Esthetic Single Tooth Sites: When Immediate, When Early, When Late?;Buser;Periodontology 2000,2017