Abstract
Two second-generation PLLA/PGA bioresorbable osteosynthetic plate systems for oral and maxillofacial surgery are available in Japan. The two systems have different PLLA-PGA component ratios (RapidSorb®, 85:15; Lactosorb®, 82:18) and plate and screw shapes. We conducted a retrospective study to compare our clinical evaluation and examine the incidence of postoperative complications between the two plate systems. A retrospective survey was conducted in 148 patients (midfacial fracture/trauma (68.2%) and dentofacial deformity patients (31.8%); males (54.7%); median age, 37.5 years) treated using maxillofacial osteosynthetic plate systems. The complications included plate exposure (7.4%), infection, (2.7%), and plate breakage (0.7%). Multivariate logistic regression analysis showed a significant correlation between sex (female), plate system (Lactosorb®), number of plates, and pyriform aperture and periorbital sites of plate placement (p < 0.05). Additionally, the propensity score-adjusted model showed a significant correlation between Lactosorb® and postoperative complications (odds ratio 1.007 (95% confidence interval, 1.001–1.055), p < 0.01). However, the two plate systems showed a low incidence rate of complications, and the plate integration and survivability were similar using 2.0-mm or 1.5-mm resorbable plate regardless of the plate system. Our findings suggest that female sex and a greater number of plates are risk factors for postoperative complications, whereas pyriform aperture and periorbital plate placements reduce the risk.
Funder
JOHNSON & JOHNSON K.K. MEDICAL COMPANY
Subject
Polymers and Plastics,General Chemistry
Cited by
18 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献