Abstract

Objective: The electrically evoked compound action potential (ECAP) is an objective measure to indirectly assess spiral ganglion neurons. The ECAP provides inputs about the prognoses of cochlear implant (CI) recipients. Several factors such as cochlear morphology can affect ECAP measurements. This study aims to investigate the variation effect of cochlear parameters on intraoperative ECAP thresholds. Methods: This is a retrospective study on patients who underwent CI surgery with normal inner ear morphology at our center between 2017 and 2023. Cochlear anatomical parameters, including diameter (A value), width (B value), and height (H value), as well as cochlear duct length (CDL), were measured pre-operatively using OTOPLAN software (Version 3.0). Cochlear implant intraoperative objective measures were also collected. The correlation between the cochlear parameters and intraoperative objective measures was studied. Results: A total of 45 patients underwent cochlear implantation. The mean age was 2.4 ± 0.9 years. The mean CDL and cochlear coverage values were 33.2 ± 2.0 mm and 76.0 ± 5.7%, respectively. The ECAP threshold increased toward basal electrodes, with ECAP values as follows: apical 13.1 ± 3.8; middle 14.3 ± 3.7; and basal 15.6 ± 4.8. Additionally, the A, B, and H values showed a positive correlation with ECAP thresholds in different cochlear regions. The B value showed a significant moderate correlation with ECAP thresholds in the middle and basal electrodes but not in the apical electrodes. Conclusions: Cochlear anatomical parameters correlate with intraoperative ECAP thresholds. The B value showed a significant association with ECAP thresholds in the middle and basal electrodes. These findings could delineate the impact of the B value in CI and optimize electrode selection. Further research is required to study this correlation and its impact on postoperative outcomes.

Publisher

MDPI AG

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3