Abstract
Air plasma spraying (APS) is a common method of producing hydroxyapatite (HA) coatings for alloprosthetic implants. Modification of HA spraying potentially may diminish the risk of inflammation and local infection during bone implantation. Titanium implants were HA coated with different deposition process parameters. HA powder was deposited using APS with axial powder injection at three different distances (100, 120, and 140 mm). The surface morphology of each implant was examined and the direct contact cytotoxicity of each HA coating was evaluated according to norm ISO 10993 5:2009. The response of monocytes to HA was assessed via the activation of transcription nuclear factor. All coatings had a lamellar structure. HA sprayed at a distance of 120 mm showed the highest roughness and little phase change. The analysis of the results of the conducted research showed that plasma-spraying distance during the HA coating process had a negligible impact on biocompatibility. The results obtained for a distance of 120 mm showed a slight increase in the biological properties tested. Moreover, HA coatings sprayed at different distances were not cytotoxic and did not stimulate the NF-kB. Bare titanium was less susceptible to colonization by Staphylococcus aureus than HA-coated surfaces. HA constitutes a potentially good, low-cost, non-cytotoxic material for joint prostheses.
Subject
Materials Chemistry,Surfaces, Coatings and Films,Surfaces and Interfaces
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献