Abstract
The actual grip force provided by a hand prosthesis is an important parameter to evaluate its efficiency. To this end, a split cylindrical handlebar embedding a single-axis load cell was designed, 3D printed and assembled. Various measurements were made to evaluate the performances of the “Federica” hand, a simple low-cost hand prosthesis. The handlebar was placed at different angular positions with respect to the hand palm, and the experimental data were processed to estimate the overall grip force. In addition, piezoresistive force sensors were applied on selected phalanxes of the prosthesis, in order to map the distribution of the grasping forces between them. The electrical current supplied to the single servomotor that actuates all the five fingers, was monitored to estimate the force exerted on the main actuator tendon, while tendon displacement was evaluated by a rotary potentiometer fixed to the servomotor shaft. The force transfer ratio of the whole system was about 12.85 %, and the mean dissipated energy for a complete cycle of closing-opening was 106.80 Nmm, resulting lower than that of many commercial prostheses. The mean grip force of the “Federica” hand was 8.80 N, that is enough to support the user in many actions of daily life, also considering the adaptive wrapping capability of the prosthesis. On average, the middle phalanges exerted the greatest grip force (2.65 N) on the handlebar, while the distal phalanges a force of 1.66 N.
Subject
Electrical and Electronic Engineering,Industrial and Manufacturing Engineering,Control and Optimization,Mechanical Engineering,Computer Science (miscellaneous),Control and Systems Engineering
Cited by
25 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献