Author:
Chen Houcai,Ge Junxiang,Kong Qingde,Zhao Zhenwei,Zhu Qinglin
Abstract
In this paper, we present the design and implementation tests of a water vapor radiometer (WVR) suitable for very long baseline interferometry (VLBI) observation. We describe the calibration method with an analysis of the sources of measurement errors. The experimental results show that the long-term measurement accuracy of the brightness temperature of the water vapor radiometer can reach 0.2 K under arbitrary ambient conditions by absolute calibration, receiver gain error calibration, and antenna feeder system temperature noise error calibration. Furthermore, we present a method for measurements of the calibration error of the oblique path measurement. This results in an oblique path wet delay measurement accuracy of the water vapor radiometer reaching 20 mm (within one month).
Funder
Astronomical Joint Fund of National Natural Science Foundation of China and Chinese Academy of Sciences
Subject
Atmospheric Science,Environmental Science (miscellaneous)
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献