Scanning of a Dental Implant with a High-Frequency Ultrasound Scanner: A Pilot Study

Author:

Bohner LaurenORCID,Habor Daniel,Radermacher Klaus,Wolfart Stefan,Marotti JulianaORCID

Abstract

The purpose of this in vitro study was to assess the trueness of a dental implant scanned using an intraoral high-frequency ultrasound prototype and compared with conventional optical scanners. An acrylic resin cast containing a dental implant at position 11 was scanned with a fringe projection 3D sensor for use as a reference dataset. The same cast was scanned 10 times for each group. Ultrasound scanning was performed with a high-frequency probe (42 MHz, aperture diameter of 4 mm and focus length of 8 mm), and 3D images were reconstructed based on the depth of each surface point echo. Optical scans were performed in a laboratory and with an intraoral scanner. A region of interest consisting of the dental implant site was segmented and matched to the reference dataset. Trueness was defined as the closeness between experimental data and the reference surface. Statistical analysis was performed with one-way ANOVA and post-hoc tests with a significance level of p = 0.05. No statistical difference was found among the evaluated scanners. The mean deviation error was 57.40 ± 17.44 µm for the ultrasound scanner, 75.40 ± 41.43 µm for the laboratory scanner and 38.55 ± 24.34 µm for the intraoral scanner. The high-frequency ultrasound scanner showed similar trueness to optical scanners for digital implant impression.

Publisher

MDPI AG

Subject

Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science

Reference37 articles.

1. Accuracy of four transfer impression techniques for dental implants: A scanning electron microscopic analysis;Yamamoto;Int. J. Oral Maxillofac. Implant.,2010

2. Accuracy of a self-perforating impression tray for dental implants

3. Accuracy of Implant Impressions for Partially and Completely Edentulous Patients: A Systematic Review

4. Evaluation of the Three-Dimensional Accuracy of Implant Impression Techniques in Two Simulated Clinical Conditions by Optical Scanning

5. Moldagem em Implantodontia;Marotti;RPG Rev. Pós Grad.,2012