Abstract
Minimum inhibition concentration (MIC) of antibiotic is an effective value to ascertain the agent and minimum dosage of inhibiting bacterial growth. However, current techniques to determine MIC are labor intensive and time-consuming, and require skilled operator and high initial concentration of bacteria. To simplify the operation and reduce the time of inhibition test, we developed a microfluidic system, containing a concentration generator and sub-micro-liter chambers, for rapid bacterial growth and inhibition test. To improve the mixing effect, a micropillar array in honeycomb-structure channels is designed, so the steady concentration gradient of amoxicillin can be generated. The flanged chambers are used to culture bacteria under the condition of continuous flow and the medium of chambers is refreshed constantly, which could supply the sufficient nutrient for bacteria growth and take away the metabolite. Based on the microfluidic platform, the bacterial growth with antibiotic inhibition on chip can be quantitatively measured and MIC can be obtained within six hours using low initial concentration of bacteria. Overall, this microfluidic platform has the potential to provide rapidness and effectiveness to screen bacteria and determine MIC of corresponding antibiotics in clinical therapies.
Funder
National Natural Science Foundation of China
State Key Laboratory of Urban Water Resource and Environment
Subject
Electrical and Electronic Engineering,Mechanical Engineering,Control and Systems Engineering
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献