Statistical Evaluation of Radiofrequency Exposure during Magnetic Resonant Imaging: Application of Whole-Body Individual Human Model and Body Motion in the Coil

Author:

Liu Wenli,Wang Hongkai,Zhang Pu,Li Chengwei,Sun Jie,Chen Zhaofeng,Xing Shengkui,Liang Ping,Wu TongningORCID

Abstract

The accurate estimation of patient’s exposure to the radiofrequency (RF) electromagnetic field of magnetic resonance imaging (MRI) significantly depends on a precise individual anatomical model. In the study, we investigated the applicability of an efficient whole-body individual modelling method for the assessment of MRI RF exposure. The individual modelling method included a deformable human model and tissue simplification techniques. Besides its remarkable efficiency, this approach utilized only a low specific absorption rate (SAR) sequence or even no MRI scan to generate the whole-body individual model. Therefore, it substantially reduced the risk of RF exposure. The dosimetric difference of the individual modelling method was evaluated using the manually segmented human models. In addition, stochastic dosimetry using a surrogate model by polynomial chaos presented SAR variability due to body misalignment and tilt in the coil, which were frequently occurred in the practical scan. In conclusion, the dosimetric equivalence of the individual models was validated by both deterministic and stochastic dosimetry. The proposed individual modelling method allowed the physicians to quantify the patient-specific SAR while the statistical results enabled them to comprehensively weigh over the exposure risk and get the benefit of imaging enhancement by using the high-intensity scanners or the high-SAR sequences.

Funder

Science Foundations of National Institute of Metrology

National Key Basic Research Project

Publisher

MDPI AG

Subject

Health, Toxicology and Mutagenesis,Public Health, Environmental and Occupational Health

Reference44 articles.

1. MRI from Picture to Proton;McRobbie,2017

2. Electromagnetic computation and modeling in MRI

3. Modelling the interaction of electromagnetic fields (10 MHz–10 GHz) with the human body: methods and applications

4. International standard, Medical Equipment IEC 60601-2-33: Particular Requirements for the Safety of Magnetic Resonance Equipment,2010

5. SAR simulations for high-field MRI: How much detail, effort, and accuracy is needed?

Cited by 10 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Polynomial chaos expansion of SAR and temperature increase variability in 3 T MRI due to stochastic input data;Physics in Medicine & Biology;2024-06-07

2. Semi-automatic Framework for Voxel Human Deformation Modeling;Current Medical Imaging Formerly Current Medical Imaging Reviews;2023-06-13

3. Evaluation of Chinese populational exposure to environmental electromagnetic field based on stochastic dosimetry and parametric human modelling;Environmental Science and Pollution Research;2023-01-07

4. Proposed Safety Guidelines for Patient Assistants in an Open MRI Environment;International Journal of Environmental Research and Public Health;2022-11-17

5. Variability in Specific Absorption Rate From Variation in Tissue Properties;IEEE Journal on Multiscale and Multiphysics Computational Techniques;2022