Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use

Author:

Lundemose Sissel Banner1ORCID,Laugesen Christian1ORCID,Ranjan Ajenthen Gayathri1,Nørgaard Kirsten1ORCID

Affiliation:

1. Steno Diabetes Center Copenhagen, Clinical Research, Diabetes Technology Research, Borgmester Ib Juuls Vej 83, DK-2730 Herlev, Denmark

Abstract

Continuous glucose monitors (CGMs) are valuable tools for improving glycemic control, yet their accuracy might be influenced by physical activity. This study sought to assess the accuracy of the three latest factory-calibrated CGM systems available in Europe at the time the study was conducted, both during aerobic exercise and in typical daily scenarios. The accuracy evaluation, based on metrics such as the median absolute relative difference (MARD) and point and rate error-grid analyses (PEGA and REGA), involved 13 adults with type 1 diabetes. Participants wore all sensors during a 1 h in-clinic exercise session followed by a subsequent 3-day home period, with blood glucose measurements serving as reference values in both contexts. During exercise, no statistically significant differences in MARD were observed (Dexcom G6: 12.6%, Guardian 4: 10.7%, and Freestyle Libre 2: 17.2%; p = 0.31), and similarly, no significant differences emerged in PEGA-zone-AB (100%, 100%, 96.8%; p = 0.37). Nevertheless, Freestyle Libre 2 showed comparatively diminished accuracy in estimating glucose trends during exercise (REGA-zone-AB: 100%, 93.0%, 73.3%; p = 0.0003). In the home environment, Freestyle Libre 2 exhibited a significantly higher MARD when compared to the other systems (10.2%, 11.9%, 16.7%, p = 0.02). Overall, Dexcom G6 and Guardian 4 demonstrated superior accuracy in both exercise and daily life scenarios compared to Freestyle Libre 2.

Publisher

MDPI AG

Subject

Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3