Wearable Electrocardiography for Physical Activity Monitoring: Definition of Validation Protocol and Automatic Classification
-
Published:2023-01-18
Issue:2
Volume:13
Page:154
-
ISSN:2079-6374
-
Container-title:Biosensors
-
language:en
-
Short-container-title:Biosensors
Author:
Cosoli Gloria1ORCID, Antognoli Luca1ORCID, Scalise Lorenzo1ORCID
Affiliation:
1. Department of Industrial Engineering and Mathematical Sciences, Università Politecnica delle Marche, v. Brecce Bianche snc, 60131 Ancona, Italy
Abstract
Wearable devices are rapidly spreading thanks to multiple advantages. Their use is expanding in several fields, from medicine to personal assessment and sport applications. At present, more and more wearable devices acquire an electrocardiographic (ECG) signal (in correspondence to the wrist), providing potentially useful information from a diagnostic point of view, particularly in sport medicine and in rehabilitation fields. They are remarkably relevant, being perceived as a common watch and, hence, considered neither intrusive nor a cause of the so-called “white coat effect”. Their validation and metrological characterization are fundamental; hence, this work aims at defining a validation protocol tested on a commercial smartwatch (Samsung Galaxy Watch3, Samsung Electronics Italia S.p.A., Milan, Italy) with respect to a gold standard device (Zephyr BioHarness 3.0, Zephyr Technology Corporation, Annapolis, MD, USA, accuracy of ±1 bpm), reporting results on 30 subjects. The metrological performance is provided, supporting final users to properly interpret the results. Moreover, machine learning and deep learning models are used to discriminate between resting and activity-related ECG signals. The results confirm the possibility of using heart rate data from wearable sensors for activity identification (best results obtained by Random Forest, with accuracy of 0.81, recall of 0.80, and precision of 0.81, even using ECG signals of limited duration, i.e., 30 s). Moreover, the effectiveness of the proposed validation protocol to evaluate measurement accuracy and precision in a wide measurement range is verified. A bias of −1 bpm and an experimental standard deviation of 11 bpm (corresponding to an experimental standard deviation of the mean of ≈0 bpm) were found for the Samsung Galaxy Watch3, indicating a good performance from a metrological point of view.
Subject
Clinical Biochemistry,General Medicine,Analytical Chemistry,Biotechnology,Instrumentation,Biomedical Engineering,Engineering (miscellaneous)
Reference62 articles.
1. Vijayan, V., Connolly, J.P., Condell, J., McKelvey, N., and Gardiner, P. (2021). Review of Wearable Devices and Data Collection Considerations for Connected Health. Sensors, 21. 2. Zhang, S., Li, Y., Zhang, S., Shahabi, F., Xia, S., Deng, Y., and Alshurafa, N. (2022). Deep Learning in Human Activity Recognition with Wearable Sensors: A Review on Advances. Sensors, 22. 3. Chen, P.-W., O’Brien, M.K., Horin, A.P., Koch, L.L.M., Lee, J.Y., Xu, S., Zee, P.C., Arora, V.M., and Jayaraman, A. (2022). Sleep Monitoring during Acute Stroke Rehabilitation: Toward Automated Measurement Using Multimodal Wireless Sensors. Sensors, 22. 4. Santucci, F., Presti, D.L., Massaroni, C., Schena, E., and Setola, R. (2022). Precordial Vibrations: A Review of Wearable Systems, Signal Processing Techniques, and Main Applications. Sensors, 22. 5. Cosoli, G., Mansi, S.A., and Arnesano, M. (2022, January 25–27). Combined use of wearable devices and Machine Learning for the measurement of thermal sensation in indoor environments. Proceedings of the 2022 IEEE International Workshop on Metrology for Living Environment (MetroLivEn), Cosenza, Italy.
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|