Abstract
Titanium (Ti) and its alloys are predominant choices for use as biomaterials in human implants. Research has shown the adverse effects of using commercial Ti alloy Ti-6Al-4V in the human body, and this presents a need for viable alternatives. In this study, Ti alloy Ti-17Nb-6Ta was manufactured by laser cladding—a prominent additive manufacturing (AM) technology. Laser cladded specimens were evaluated for their in vitro and electrochemical behavior. A human osteosarcoma cell line (MG-63 cells) was used for in vitro investigations. Cell proliferation was good in the physiological medium, and cells were alive when in contact with the laser cladded alloy, even after two to three weeks, indicating good cell viability and compatibility with this alloy. Electrochemical characterization was carried out in Ringer’s solution, and noticeably lower corrosion current density and corrosion rate values were observed. The lower amounts of these parameters indicated the passivation behavior due to multi-layer Ti, Nb, and Ta alloy oxide films. These oxide films also enhanced osseointegration. Thus, the Ti-17Nb-6Ta alloy can be an ideal biocompatible alternative to Ti-6Al-4V.
Subject
Inorganic Chemistry,Condensed Matter Physics,General Materials Science,General Chemical Engineering
Cited by
3 articles.
订阅此论文施引文献