Abstract
To evaluate the automated determination of the center of an idiopathic macular hole (MH) by using swept-source optical coherence tomography (OCT) images with new macro-based algorithms in ImageJ and to compare the difference between the MH center measurements obtained automatically and manually. This cross-sectional study included 39 eyes of 39 elderly individuals (22 women, 17 men) with stage 3 and 4 MH. The MH center was automatically determined using the ImageJ macro. The foveal center was also manually identified by two masked examiners using horizontal and vertical serial B-scan OCT angiography images. The mean age was 68.8 ± 8.3 years. After adjusting for the effect of magnification, the mean distance between the MH center determined manually by Examiner 1 and that determined automatically was 15.5 ± 9.9 µm. The mean distance between the two manually determined measurements of the MH center was 20.3 ± 19.7 µm. These two mean distance values did not differ significantly (Welch t-test, p = 0.27) and was non-inferior (p < 0.0001). The automated ImageJ-based method for determining the MH center was comparable to manual methods. This study showed that automated measurements were non-inferior to manual measurements, and demonstrated a substitutable usefulness, at least for use in clinical practice.