Use of Thermoluminescence Dosimetry for QA in High-Dose-Rate Skin Surface Brachytherapy with Custom-Flap Applicator

Author:

Manna Francesco12,Pugliese Mariagabriella13ORCID,Buonanno Francesca4ORCID,Gherardi Federica4,Iannacone Eva4ORCID,La Verde Giuseppe13ORCID,Muto Paolo4,Arrichiello Cecilia4

Affiliation:

1. Department of Physics “E. Pancini”, Federico II University, 80126 Naples, Italy

2. Centro Servizi Metrologici e Tecnologici Avanzati, Federico II University, 80146 Naples, Italy

3. National Institute of Nuclear Physics, Section of Naples, 80126 Naples, Italy

4. Radiotherapy Unit, Istituto Nazionale Tumori, Istituto di Ricovero e Cura a Carattere Scientifico (IRCCS), Fondazione G. Pascale, 80131 Naples, Italy

Abstract

Surface brachytherapy (BT) lacks standard quality assurance (QA) protocols. Commercially available treatment planning systems (TPSs) are based on a dose calculation formalism that assumes the patient is made of water, resulting in potential deviations between planned and delivered doses. Here, a method for treatment plan verification for skin surface BT is reported. Chips of thermoluminescent dosimeters (TLDs) were used for dose point measurements. High-dose-rate treatments were simulated and delivered through a custom-flap applicator provided with four fixed catheters to guide the Iridium-192 (Ir-192) source by way of a remote afterloading system. A flat water-equivalent phantom was used to simulate patient skin. Elekta TPS Oncentra Brachy was used for planning. TLDs were calibrated to Ir-192 through an indirect method of linear interpolation between calibration factors (CFs) measured for 250 kV X-rays, Cesium-137, and Cobalt-60. Subsequently, plans were designed and delivered to test the reproducibility of the irradiation set-up and to make comparisons between planned and delivered dose. The obtained CF for Ir-192 was (4.96 ± 0.25) μC/Gy. Deviations between measured and TPS calculated doses for multi-catheter treatment configuration ranged from −8.4% to 13.3% with an average of 0.6%. TLDs could be included in clinical practice for QA in skin BT with a customized flap applicator.

Publisher

MDPI AG

Subject

Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry

Reference75 articles.

1. GEC-ESTRO ACROP recommendations in skin brachytherapy;Guinot;Radiother. Oncol.,2018

2. Current status of brachytherapy in cancer treatment—Short overview;Skowronek;J. Contemp. Brachyther.,2017

3. Treatment planning methods in high dose rate interstitial brachytherapy of carcinoma cervix: A dosimetric and radiobiological analysis;Anbumani;ISRN Oncol.,2014

4. Pulsed brachytherapy: The conditions for no significant loss of therapeutic ratio compared with traditional low dose rate brachytherapy;Fowler;Int. J. Radiat. Oncol. Biol. Phys.,1992

5. International Commission on Radiation Units and Measurements (1985). Report 38 of ICRU. Dose and Volume Specification for Reporting Intracavitary Therapy in Gynecology, ICRU.

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3