Abstract

The most commonly used material in dental implants and their abutments is titanium. Zirconia is a more aesthetic alternative to titanium abutments; however, it is much harder. There are concerns that zirconia could damage the surface of the implant over time, especially in less stable connections. The aim was to evaluate the wear of implants with different platforms connected to titanium and zirconia abutments. A total of six implants were evaluated, two of each connection type: external hexagon, tri-channel, and conical connections (n = 2). Half of the implants were connected to zirconia abutments, and the other half to titanium abutments (n = 3). The implants were then cyclically loaded. The implant platforms were evaluated by digital superimposing micro CT files and calculating the area of the loss surface (wear). In all the implants, a statistically significant loss of the surface area (p = 0.028) was observed when comparing the area before and after cyclic loading. The average lost surface area was 0.38 mm2 with titanium abutments and 0.41 mm2 with zirconia abutments. The average lost surface area was 0.41 mm2 with the external hexagon, 0.38 mm2 with the tri-channel, and 0.40 mm2 with the conical connection. In conclusion, the cyclic loads induced implant wear. However, neither the type of abutment (p = 0.700) nor the connection (p = 0.718) influenced the amount of surface area lost.

Funder

FCT–Fundação para a Ciência e a Tecnologia

Publisher

MDPI AG

Subject

Biomedical Engineering,Biomaterials

Reference43 articles.

1. A Brief Historical Perspective on Dental Implants, Their Surface Coatings and Treatments;Abraham;Open Dent. J.,2014

2. Biomedical applications of titanium and its alloys;Elias;Jom,2008

3. Implant materials, designs, and surface topographies: Their effect on osseointegration. A literature review;Sykaras;Int. J. Oral Maxillofac. Implant.,2000

4. Aluminum, Alzheimer’s disease and bone fragility;Mjoberg;Acta Orthop. Scand.,1997

5. Accumulation of aluminum in lamellar bone after implantation of titanium plates, Ti-6Al-4V screws, hydroxyapatite granules;Zaffe;Biomaterials,2004

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3