Do Titanium Mini-Implants Have the Same Quality of Finishing and Degree of Contamination before and after Different Manipulations? An In Vitro Study

Author:

Zogheib Tatiana,Walter-Solana André,de la Iglesia Fernando,Espinar Eduardo,Gil JavierORCID,Puigdollers Andreu

Abstract

Evaluate the quality of finishing and degree of contamination before and after handling and surface treatment of titanium (Ti) orthodontic mini-implants (OMIs). A scanning electron microscope (SEM) study on ninety-six titanium OMIs was done. Energy-Dispersive X-ray Analysis (EDX) identified the present particles on manufactured OMIs surfaces. Then, OMIs were manipulated with gauze (dry sterile, soaked in chlorhexidine) and gloves (latex, nitrile) to evaluate the contamination of these handling materials. Finally, OMIs underwent surface treatments and were placed in bone to observe the contaminants they released. Roughness (Ra) and wettability with contact angle parameter (CA) were measured on these treated OMIs. Machined OMIs presented surface irregularities and were contaminated with manufacturing-process particles (carbon, plastic Polyvinyl Chloride PVC, aluminum). Hand-manipulated OMIs were also contaminated by the handling materials. OMIs surface characteristics were as follows: acid-etched (Ra ≈ 1.3 μm, CA ≈ 66°), machined (Ra ≈ 0.3 μm, CA ≈ 68°), SB (Ra ≈ 3.3 μm, CA ≈ 78°), and SBAO (Ra ≈ 3.1 μm, CA ≈ 92°). Bone was contaminated by OMIs surface defects and extra particles. Manufactured OMIs have surface contaminants that increase with clinical handling. Surface treatments (SBAO, a combination of sandblasting and anodic oxidation) increase the roughness and contact angle, which play an important role in osseointegration. Surface-treated OMIs leave titanium particles in the bone during their insertion-removal. The use of a gauze soaked in chlorhexidine is recommended when handling OMIs. Further investigations would be interesting to study more variables and confirm the present results.

Publisher

MDPI AG

Subject

General Materials Science,Metals and Alloys

Reference56 articles.

1. Miniscrews as orthodontic anchorage: A preliminary report;Costa;Int. J. Adult Orthod. Orthognath. Surg.,1998

2. Mini-implant for orthodontic anchorage;Kanomi;J. Clin. Orthod. JCO,1997

3. The skeletal cortical anchorage using titanium microscrew implants;Park;Korea J. Orthod.,1999

4. Mini-implant anchorage for the orthodontic practitioner

5. Root proximity is a major factor for screw failure in orthodontic anchorage

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3